CGRP receptor bound to CGRP dataset Sample: RAMP1-CLR-CGRP (CGRPR-CGRP) Grid: UltrAuFoil 300 mesh R1.2/1.3 Cemera: Gatan K3 post a BioQuantum energy filter Camera mode: CDS, counting (hardware binning by 2) TEM mode: EFTEM NanoProbe Mag: x105k C2 aperture: 50 um Spot: 5 Beam diameter: 1.6 um EF slit: 25 eV Data collection software: SerialEM Acquisition method: beam-image shift with beam-tilt compensation Acquisition pattern: 3 x 3 holes, 1 image in the center of each hole Physical Pixel: 0.83 A/pix Exposure rate: 7.2 e/pix/s (CDS) Exposure time: 5.011 s Total exposure: 52.4 e/A2 Frames: 71 Defocus: 0.7 - 1.6 um Movies: 5598 Data format: non-gain-normalized LZW compressed TIFFs Gain reference: "RAMP1-CGRP_gain.mrc" prepared from the data Final particles: 388k Particles file: "CGRPR-CGRP_data.star" Optics groups: per-acquisition position Final resolution: 3.5 A Rado March 19, 2021