The dataset was collected on JEOL CRYO ARM 200 II microscope at 200 kV accelerating voltage, Cs = 1.53 mm, Cc = 1.8 mm. Folder contents: 20231222_231005.apoF-GFP.4 - non-gain-normalized movies in compressed TIFF format auxiliary_files - gain reference "231005.apoF-GFP.4_gain.mrc"; CryoSPARC processing workflow "200kv_apoferritin_workflow.json"; Exported CryoSPARC particle stack from the last 3D refinement job "J103_particles_exported.cs" Related EMDB and PDB entries: EMD-68251 PDB-22FX Experiment details: ○ Mag.: x150k ○ Pixel size: 0.3056 A/physpix ○ CLA 100 um ○ Spot 3 ○ Angle 3 ○ Focus area offset 1.2 um ○ View mode: x5k, Spot 7, Angle 8, CL3: C0B3 ○ Beam diameter ~950 nm ○ NO OLA ○ C3: 8AA8 ○ BA2: 8410 7813 ○ Exp. rate: 3.3 e/physpix/s ○ Dose rate: 35.335 e/A2/s ○ Exposure time: 1.51 s ○ Total dose: 53.356 e/A2 ○ Frame time: 0.0185 s ○ Frames: 81 ○ Exposure per frame: 0.6587 e/A2/frame ○ BS-IS acquisition pattern: 3x3x1 in the center of each hole ○ Data was collected in non-gain-normalized "Counting" mode with CDS, e.g. with 2x hardware binning. ○ Movie size ~125 MB ○ Target defocus -0.5 um with 0.02 um increments between holes. ○ Total number of micrographs: 13,654 collected in ~56 h; average throughput: ~244 movies/hour.